| Discipline | Parameter | Range | Best Uncertainty | Method |
|---|---|---|---|---|
| Electrical | DC Voltage | 0 mV – 1000 V | 12 ppm | Direct comparison against reference multimeter |
| Electrical | DC Current | 0 µA – 20 A | 35 ppm | Shunt comparison |
| Electrical | AC Voltage (45 Hz – 1 kHz) | 1 mV – 1000 V | 120 ppm | Thermal transfer standard |
| Electrical | Resistance | 1 mΩ – 100 MΩ | 18 ppm | Four-wire comparison |
| High Resistance and Insulation | Insulation Resistance | 100 kΩ – 10 TΩ | 0.8 % | High resistance standard comparison |
| Temperature and Humidity | Temperature (liquid bath) | −30 °C to +250 °C | 0.03 °C | Comparison against SPRT |
| Temperature and Humidity | Temperature (dry block) | +250 °C to +1200 °C | 0.45 °C | Comparison against type R thermocouple |
| Temperature and Humidity | Relative Humidity | 10 % – 95 % RH | 0.8 % RH | Two-pressure humidity generator |
| Pressure | Pneumatic Pressure | −95 kPa to 7 MPa | 0.010 % of reading | Pressure balance comparison |
| Pressure | Hydraulic Pressure | 0 – 100 MPa | 0.018 % of reading | Deadweight tester |
| Torque | Static Torque | 0.5 – 1000 N·m | 0.25 % | Reference transducer comparison |
| Acoustic and Sound | Sound Level Meter | 94 dB / 114 dB at 1 kHz | 0.15 dB | Acoustic calibrator comparison |
| Speed and Vibration | Rotational Speed | 10 – 100,000 rpm | 0.05 % | Optical reference comparison |
| Speed and Vibration | Vibration Acceleration | 10 Hz – 10 kHz | 1.2 % | Back-to-back accelerometer comparison |
| Gas Measurement | Combustible Gas (% LEL) | 0 – 100 % LEL | 2.0 % of reading | Certified reference gas comparison |
| Gas Measurement | Toxic Gas (CO, H₂S) | 0 – 500 ppm | 2.5 % of reading | Certified reference gas comparison |